December 2016, Volume 66, Issue 12

Original Article

Impact of brachytherapy applicators geometry on dose distribution in gynaecological cancer

Ateeque Ur Rehman  ( Department of Physics, The Islamia University of Bahawalpur, Pakistan. )
Saeed Ahmad Buzdar  ( Department of Physics, The Islamia University of Bahawalpur, Pakistan. )
Muhammad Hassan  ( Department of Physics, The Islamia University of Bahawalpur, Pakistan. )
Asghar Gadi  ( Bahawalpur Institute of Nuclear Medicine and Oncology, Pakistan. )
Muhammad Afzal Khan  ( Department of Physics, The Islamia University of Bahawalpur, Pakistan. )
Muhammad Naeem Anjum  ( Department of Physics, The Islamia University of Bahawalpur, Pakistan. )

Abstract

Objective: To evaluate the efficacy of suitable applicators for intracavitary brachytherapy to treat cervical cancer.
Methods: This study was conducted at the Bahawalpur Institute of Nuclear Medicine and Oncology, Bahawalpur, Pakistan, in February 2014. Dose distributions for Fletcher and ring applicators were analysed at target points according to recommended protocols using Abacus software. The doses to the organs at risk, i.e. bladder and rectum, were also focused. SPSS 16 was used for data analysis.
Results: A total of 40 patients\\\' plans were considered. The mean absorbed dose at target point A was 6.87±0.201Gy in case of Fletcher applicator (p=0.082) and 6.79±0.107Gy in case of ring applicator (p=0.001). Absorbed dose at the target point was significantly higher for the Fletcher type applicators as compared to the ring applicators. However, the amount of doses at bladder and rectum were smaller for the Fletcher applicators.
Conclusion: Fletcher applicator was more effective in achieving a better dose distribution in gynaecological malignancies, resulting in better treatment outcome.
Keywords: Intracavitary Brachytherapy, Applicator Geometry, Dose, Gynaecological Cancer. (JPMA 66: 1566; 2016)


Introduction
The goal of radiotherapy is to deliver the maximum dose to the malignant tissues while minimising the exposure of surrounding healthy tissues. Nowadays radiotherapy alone or the combination with chemotherapy, brachytherapy or surgery is the major treatment technique for the cancer.1
Brachytherapy is an integral component in any radiotherapy department. Intracavitory brachytherapy (ICBT) is the most commonly performed procedure, while interstitial brachytherapy is conserved for selected indication, e.g. narrow vagina or where there is no body cavity.2 Therefore, different geometry of the applicators plays a crucial role in the ICBT.3
The subject of considerable research is to analyse the dose distribution scheme to the target by using Fletcher-style applicator and ring-type applicator. Various recommendations are available for intracavitary insertion technique, dosage schedule, dose prescription as well as for reporting of the full ICBT treatment procedure.4


Choice of the ICBT applicator is rather arbitrary and also depends upon the accessibility of the applicator type.5 The dose distribution in the brachytherapy is essentially dependent on the inverse square law; so different types of dose distribution are attained with two different applicator systems, which may help to achieve higher therapeutic ratio.6 The anatomy of the patient and the extension of the disease also play a crucial role in the selection of a suitable applicator for the treatment. Different types of the applicators are being used to treat the cancer in brachytherapy. The size and specification of the applicators are allied to the dose delivered to adjacent organs.7 Radioisotopes are directly placed into the tumour or placed in the applicator which has been initially inserted into the body cavity close to the tumour.8 International working party for the early diagnosis and treatment of cervical cancer was introduced in 1972 by the co-operation of the International Atomic Energy Agency (IAEA) and the World Health Organisation (WHO) in developing countries.9


Assembly of the applicators depends upon the gynaecological malignant part. For example, vaginal treatment is instigated by the vaginal cylinder, and the cervical, uterine or endometrial cancer is initiated by the tandem and ovoid or tandem and ring applicator of various sizes. Tandem and ovoid devices are being used to treat gynaecological cancer such as uterine, cervical malignancies, etc. Numerous rigid forms of Fletcher, Henschke and ring applicators are also being used.10
In developed countries, image-guided brachytherapy (IGBT) is being used in many oncology centres, but in Pakistan both two- and three-dimensional treatment planning systems are used. The current study was planned to evaluate the dose distribution scheme of these different shape applicators to enhance the quality of treatment.


Materials and Methods
This dosimetric study was conducted in February 2014 at the Bahawalpur Institute of Nuclear Medicine and Oncology (BINO), Bahawalpur, Pakistan, where Fletcher and ring applicators are used for cervical cancer.
The Fletcher applicator was introduced in the early 1950s for treating carcinoma of the cervix. With the passage of time, some modifications have been made in the applicator to increase its suitability and to transfer the maximum dose to the target.11 In this research work, Titanium Fletcher applicator of 6cm tandem length and intrauterine angle of 45° was used. The accessible Fletcher-style applicator (Varian Medical Systems) consists of two cylindrical vaginal ovoids and intrauterine tube. The ovoids lie at a plane almost perpendicular to the plane of the uterine tandem. Source stopping position (SSP) is kept at 0.5cm.
Ring applicator (Varian Medical Systems) consists of the uterine tandem and a vaginal ring. The ring and intrauterine tube are fixed to each other by a screw. The cap on the ring tube reduces the dose to vaginal mucosa. Fixed geometry is the advantage of ring applicator but curvature of the ring may cause difficulty for high dose rate (HDR) source to navigate.12 In the current study, ring applicator of 6cm tandem length and angle of 45° was used, whereas ring diameter was 3.2 cm and SSP was kept at 0.5cm.


Treatment plans were generated using Abacus 3.0 treatment planning system (Varian Medical Systems) by digitising the dose reference point (DRP) and SSP for each plan. To perceive the suitability of both applicators, different optimising tools were applied. The basic parameters of the applicators for insertion like intrauterine, tandem length and angulations 45° were kept similar for both applicators. All suitable considerations were used to maximise the dose to the target point A and to minimise at rectum, bladder and vaginal mucosa. After the insertion of applicator, anteroposterior and lateral (AP/LAT) view of digitally reconstructed radiographs (DRRs) were taken to content the rectum and bladder (organs at risk, or OARs).


The position of bladder and rectum was located and then reconstructed on the radiograph on Abacus treatment planning system. According to the Manchester system\\\'s five reference points — targetpoint A (left and right), R for the rectum point, and B1 and B2 for the bladder — were reconstructed and dose was prescribed according to International Commission on Radiation Units and Measurements report (ICRU 38).13,14 Point A is defined as 2cm in the superior beside the tandem from the flange and 2cm lateral to the uterine duct. 15 The dose fraction at point A, rectum and bladder was 7.00Gy, 5.00Gy, and 5.60Gy, respectively. Optimising constraints were used to deliver the prescribed dose to the point A and to secure the OARs. Approximately pear shape isodose curves were obtained for both applicators.


Dose distribution scheme of isodose curves was observed to compare both applicators. SPSS 16 was used to calculate the mean absorbed dose and standard deviation. T-test was used to find out the significance of the difference of absorbed dose. Absorbed dose to the target point was noted. Rectal and bladder dose was also observed to check out the suitability of the applicators. To check the level of significance, p-value was calculated with confidence interval (CI) at 95%. For all insertions, ICRU report 38 and American Brachytherapy Society (ABS) recommendations were kept in mind.14-16


Results
A total of 40 patients\\\' plans were considered for this study. The mean absorbed dose at target point A was 6.87±0.201Gy in case of Fletcher applicator (p=0.082) and 6.79±0.107Gy in case of ring applicator (p=0.001). The respective values were 4.45±0.705Gy (p=0.038) and 4.80±0.410Gy (p=0.170) for rectum point, and 5.82±0.360Gy (p=0.085) and 5.89±0.181Gy (p=0.001) for bladder point (Table)


AP/LAT views of isodose curves showed that absorbed dose at target point was slightly different for both applicators but there was no statistically significant difference noted (Figure-1)


Each plan number contained an average of four treatment plans. This represented 7.00 Gy was prescribed to the target point A, but slightly different absorbed dose was recorded in both Fletcher and ring cases. Suitable dose was transferred to the target point in case of Fletcher compared to the ring applicator (Figure-2a)


Moreover, 5.60Gy was the limited dose for the bladder, but in some cases for both applicators slightly higher or lesser dose was transferred to the bladder. To chase the aim of every treatment planning, minimum dose should be delivered to the OARs, so according to the graphical representation less dose was transferred to the bladder in case of Fletcher as compared to the ring applicator (Figure-2b).
Furthermore, 5.00Gy was the limited dose for the rectum, but slightly higher dose was shifted to the rectum in case of ring applicator as compared to the Fletcher (Figure-2c).


Discussion
Geometry of the applicators plays a vital role in the dose distribution in brachytherapy.6 Both ring and Fletcher applicators are based on the Manchester system but their geometry is quite different. Curved shape of the ring applicator differs the scheme of dose distribution of both applicators. 17 When these applicators are used in the same patient, they might show the different isodose curves. In this study, dose at the target point A and ICRU reference points (bladder and rectum) were noted. To compare the dose distribution around the OARs, isodose curves were also observed. According to the ICRU recommendation, dose was prescribed to the reference point and OARs were secured because these high-energy radiations produce the hydroxyl ion and damage the normal tissues.18
Sukhvir Singh et al. studied the effects of geometry of fixed- and flexible-shape applicators by using Abacus treatment planning software under the ICRU report 38 recommendations. They observed the dose variation at OARs (rectum and bladder). According to their results, in case of flexible geometry higher dose was delivered as compared to the fixed geometry applicator and isodose thickness was also higher in fixed geometry.6 Bishan Basu et al. compared the Manchester and Fletcher applicators. They used the dose volume data for the comparison. According to their results, Manchester applicator covered wider target volume as compared to the Fletcher.19 Tuncell et al. compared the ring and ovoid dose distribution to study the post-operative vaginal cuff irradiation, therefore no tandem was used. In their study, less dose was delivered to the rectum using ring as compared to ovoid; but the drawback of their study was that due to the absence of tandem, vaginal packing was not so better.20


In the current study, the mean absorbed dose of all the patients was 6.79±0.107Gy (p=0.001) in case of ring applicator and 6.87±02.01Gy (p=0.082) in case of Fletcher applicator. Our findings showed that Fletcher readings were more reliable than ring applicator. For the comparison of bladder dose, the mean absorbed dose for ring applicator was 5.89±0.181Gy (p=0.001) and 5.82±0.360Gy (p=0.08) for Fletcher applicator. In case of ring applicator higher dose was delivered to the bladder than Fletcher. Besides, average rectal dose of 4.80±0.410Gy dose was delivered to the rectum in case of ring applicator (p=0.17) and 4.45±0.705Gy in case of Fletcher applicator (p=0.038).


In case of Fletcher applicator, dose value at the target point was insignificantly less (p= 0.08) or approximately equal to the prescribed dose, but in case of ring applicator absorbed dose was significantly less (p=0.001) than the prescribed dose value. To cure the OARs, minimum dose should be delivered to the rectum and bladder. Similarly for the rectum both applicators delivered less dose as compared to the prescribed dose, but in case of Fletcher applicator significantly less (p=0.038) dose was delivered to the rectum. On the other hand, both applicators delivered slightly higher dose to the bladder point than the prescribed dose, but ring applicator delivered significantly higher dose. There was no significant difference in the mean absorbed dose at point A for both applicators. Slightly higher dose was delivered in case of Fletcher applicator at target point A than ring applicator. The dose received by the OARs, e.g. the urinary bladder and the rectum, was different too. The dose received by the urinary bladder and rectum was significantly higher when ring applicator was used.


Conclusion
Absorbed dose at the target point was significantly higher for the Fletcher-type applicators as compared to the ring applicators. However, the amount of doses at bladder and rectum was less for the Fletcher applicators. Therefore, Fletcher applicator achieved a better dose distribution in gynaecological malignancies, which resulted in better treatment outcome.


Disclaimer: None.
Conflict of Interest: None.
Source of Funding: None.


References
1.Deep K, Anand V, Shrikant A. Nanoparticles in radiation therapy: a summary of various approaches to enhance radiosensitization in cancer. Transl Cancer Res 2013; 2: 330-42.
2.Hsiang CK, Keyur JM, Ravindra Y MS, Linda H, Dinesh M, et al. Feasibility study and optimum loading pattern of a multi-ring inflatable intravaginal applicator. J Contemp Brachytherapy. 2013; 5:93-100.
3.Hsu IC, Speight J, Hai J, Vigneault E, Phillips T, Pouliot J. A comparison between tandem and ovoids and interstitial gynecologic template brachytherapy dosimetry using a hypothetical computer model. Int J Radiat Oncol Biol Phys. 2002; 52:538-43.
4.Management of cervical cancer: Strategies for limited resource centers. A guide for radiation oncologists. International Atomic Energy Agency; Vienna, 2013.
5.Basu B, Basu S, Chakraborti B, Ghorai S, Gupta P, Ghosh S. A comparison of dose distribution from Manchester-style and Fletcher-style intracavitary brachytherapy applicator systems in cervical cancer. J Contemp Brachytherapy. 2012; 4:213.
6.Sukhvir S, Arti S, Manoj KS, Ajay KS, Pankaj G, Ashok K. Dosimetric Comparison of Fixed and Flexible Geometry ICRT Applicators in Patients Treated for Carcinoma Cervix. IJMPCERO. 2015;4:1-6.
7.Sharma DN, Chaudhari P, Sharma S, Gupta L, Jagadesan P, Rath GK, et al. Comparison of high-dose-rate intracavitary brachytherapy dosimetry with and without anesthesia in patients with cervical carcinoma. J Applied Clinical Medical Physics. 2014; 15:60-6.
8.Cuaron JJ, Hirsch JA, Medich DC, Rosenstein BS, Martel CB, Hirsch AE. A Proposed Methodology to Select Radioisotopes for Use in Radionuclide Therapy.AJNR Am J Neuroradiol. 2009; 30:1824-9.
9.Calibration of BrachytherapySources.Vienna, Austria:International Atomic EnergyAgency;1999.
10.William MR, Sudershan KB, Yusung K. The dosimetric impact of vaginal balloonpacking on intracavitary high-dose-rate brachytherapy for gynecological cancer. J Contemp Brachytherapy. 2013; 5:17-22.
11.Haas JS, Dean RD, Mansfield CM. Dosimetric comparison of the Fletcher family of gynecologic colpostats 1950-1980.Int J Radiat Oncol Biol Phys.1985; 11: 1317-21.
12.McMahon R, Zhuang T, Steffey BA, Song H, Craciunescu OI. Commissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets. J Appl Clin Med Phys. 2011; 12:3447.
13.Nag S, Erickson B, Thomadsen B, Orton C, Demanes JD, Petereit D. The American Brachytherapy Society recommendations for high-dose-rate brachytherapy for carcinoma of the cervix. Int J Radiat Oncol Biol Phys. 2000; 48:201-11.
14.ICRU Report 38: Dose and Volume Specification for Reporting Intracavitary Gynecology, 1985.
15.Khan FM. The physics of Radiation Therapy. 4th edition. Philadelphia, Pa: Lippincott Willians & Wilkins; 2010.
16.Viswanathan AN, Thomadsen B, American Brachytherapy Society Cervical Cancer Recommendations Committee; American Brachytherapy Society. Consensus guidelines for locally advanced carcinoma of the cervix. Part I: general principles. Brachytherapy. 2012; 11: 33-46.
17.McMahon R, Zhuang T, Steffey BA, Song H, Craciunescu OI. Commissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets. J Appl Clin Med Phys. 2011; 12:3447.
18.Deborah C, Ana PC, James BM. Radioprotectors and Mitigators of Radiation- induced normal tissue Injury. Oncologist. 2010; 15:360-71.
19.Basu B, Basu S, Chakraborti B, Ghorai S, Gupta P, Ghosh S, et al. A Comparison of dose distribution from Manchester-style and Fletcher-style intracavitary brachytherapy applicator systems in cervical cancer. J Contemp Brachytherapy. 2012; 4:213-8.
20.Tuncell N, Toy A, Demiral AN, Cetingoz R, Ganpaqaoglu M. Dosimetric comparison of ring and ovoid applicators. J BUON. 2009; 14:451-5.

Journal of the Pakistan Medical Association has agreed to receive and publish manuscripts in accordance with the principles of the following committees: